Thermal Desorption Tube, Anasorb GCB1/Carbosieve S-III, Conditioned, 1/4in OD x 3 1/2in size stainless steel tube, 2 sections, 280/165 mg sorbent, 60/80 mesh

Part Number:

226-349

Limited Shelf Life
Made to Order
- +
Overview
Thermal Desorption Tube, Anasorb® GCB1/Carbosieve S-III, 2 sections, 250/150 mg, 60/80 mesh, in a 1/4-inch OD x 3 1/2-inch size stainless steel tube. Tube Ends: Stainless Steel Open. Separators: Glass Wool/Glass Wool/Glass Wool/Screen. Meets EPA Method TO-17 and is compatible with Perkin Elmer and Markes International thermal desorbers. Each tube is thermally conditioned before shipping, identified with a unique serial number and flow direction arrow, and should be used within six months or reconditioned before use. Restocking fee may apply. Also available unconditioned as Cat. No. 226-349-UP.

At a Glance

SKC Thermal Desorption Tubes are designed for active sampling of ppb-level volatile organic compounds (VOCs) in indoor and ambient air. Available with a wide selection of high-performance sorbents and single- or multiple-bed configurations, these reusable tubes support EPA, ASTM, MDHS, and NIOSH thermal desorption method requirements. Offered in conditioned (ready-to-use) and unconditioned options, SKC Thermal Desorption Tubes are compatible with PerkinElmer and Markes International thermal desorbers and provide reliable VOC collection with high analytical sensitivity and recovery.

 

Key Features and Benefits

Compatible with thermal desorbers: Designed for analysis using thermal desorption techniques. Tubes are compatible with PerkinElmer and Markes International thermal desorbers.

Multiple sorbent options: Available in single and multiple-bed sorbent configurations. Choose from a range of sorbent materials for different applications. See table below.

Reliable sample collection: Sorbent materials are selected to effectively collect and retain target compounds.

Low background: Quality stainless steel and glass tube materials and sorbents help minimize background contributions during analysis.

Reusable: Tubes can be conditioned and reused when appropriate, providing an economical sampling option.

Broad applications: Suitable for occupational, environmental, indoor air, and other air sampling applications.

Supports EPA, ASTM, MDHS, and NIOSH thermal desorption methods.

 

Available Sorbents

SKC Thermal Desorption Tubes are available in single and multiple-bed configurations with a variety of sorbents to support a wide range of sampling applications.

Anasorb™ 747 (20/40 mesh size and 980 m2/gm surface area) 

This beaded active carbon has the capacity for organic vapors similar to petroleum-based and coconut-shell charcoals. It can also be used to sample non-polar organic compounds.

Carbosieve S-III (60/80 mesh size and 975 m2/gm surface area) 

This carbon-based molecular sieve has properties similar to the obsolete Anasorb CMS and replaces it in SKC Thermal Desorption Tubes. Carbosieve S-III has a high capacity/breakthrough volume for small molecules and is moderately hydrophilic. It can be used for lower-boiling compounds such as halogenated hydrocarbons and alcohols.

Anasorb GCB1 (20/40 or 60/80 mesh size and 100 m2/gm surface area) 

Anasorb Graphitized Carbon Black 1 (GCB1) has a moderate surface area and is equivalent to Carbopack B (60/80). This sorbent has proven to be valuable for sampling compounds of intermediate to high volatility.

Chromosorb® 106 (60/80 mesh size and 700 to 800 m2/gm surface area) 

A porous polymer sorbent, Chromosorb 106 is used to trap low-boiling hydrocarbons, benzene, labile compounds, and volatile oxygenated compounds. This hydrophobic sorbent is the least polar polymer in the Chromosorb family and is functionally similar to Anasorb 727.

Anasorb GCB2 (20-40 or 60/80 mesh size and 10 m2/gm surface area) 

This sorbent has a low surface area and is equivalent to Carbopack C (60/80). Anasorb GCB2 is useful in collecting semi-volatile organic compounds (SVOCs).

Tenax® TA (35/60 or 60/80 mesh size and 30 to 35 m2/gm surface area) 

Tenax TA is a traditional sorbent (porous polymer) for trapping medium to high-boiling compounds; it is especially useful for low concentrations because of its low background. Tenax TA is hydrophobic and is suitable for use in EPA Method TO-17 or IP-1B and other thermal desorption applications.

Carbopack X (40/60 mesh size and 240 m2/gm surface area) 

 

This strong graphitized carbon sorbent collects C3 to C9 hydrocarbons. Carbopack X is hydrophobic.

Tenax GR (60/80 mesh size and 35 m2/gm surface area) 

This Tenax/graphite composite extends the range of Tenax to lower-boiling compounds. It retains Tenax's ability to be cleaned easily to a very low background and is widely used in thermal desorption applications.


Applications

  • Indoor air quality monitoring
  • Ambient air monitoring
  • Occupational exposure monitoring
  • Volatile organic compound (VOC) sampling
  • Environmental air sampling
  • Workplace air monitoring
  • Method-specific EPA, ASTM, MDHS, and NIOSH air sampling

Why Choose SKC Thermal Desorption Tubes

SKC Thermal Desorption Tubes combine advanced sorbent technology with high quality materials for reliable collection of volatile organic compounds (VOCs) at ppb-level concentrations. Engineered with low background, thermally stable sorbents that provide high analytical recovery, these reusable tubes support sensitive VOC analysis without the sample dilution associated with solvent extraction. Available in a variety of sorbent configurations, SKC Thermal Desorption Tubes provide dependable performance for occupational, indoor, and ambient air monitoring applications.

Attributes
Cold Storage RecommendedNo
Limited Shelf LifeYes
Made to OrderYes
TypeThermal Desorption
SorbentAnasorb GCB1, Carbosieve S-III
Sorbent TreatmentConditioned
Number of Sections2
Tube MaterialStainless Steel
Tube Cover SizeB (7 or 8-mm x 110mm)