UMEx 400 Passive Sampler for Aliphatic Amines, 10Pk

Part Number:

500-400

Cold storage required. Shipping restrictions may apply.
Limited Shelf Life
- +
Overview
UMEx 400 Passive Sampler, for aliphatic amines, with a validated concentration range of 0.5 to 10 ppm; single use only.
The SKC UMEx 400 Passive Sampler contains a tape treated with 1-Naphthyl-isothiocyanate (NITC) to collect aliphatic amines in the 0.5 to 10 ppm concentration range. Analysis is by high-performance liquid chromatography with UV detection (HPLC-UV) for identification of amines that may be present in the sample. Designed for single use, the SKC UMEx 400 Passive Sampler is packaged in an aluminized pouch for easy sample transport to a laboratory. 
Attributes
Cold Storage RecommendedYes
Limited Shelf LifeYes
TypeSolvent Extraction
Chemical HazardAliphatic Amines
SorbentTreated Tape
Validation LevelPartial

Performance Profile

Sampling Rates:
CompoundSampling Rate (ml/min)RSD
Methyl amine18.414.0
Dimethyl amine18.25.8
Isopropyl amine13.99.5
n-Butyl amine18.19.2

 

Concentration Range:0.5 to 10 ppm
Detection Principle:Formation of a stable amine derivative, made from a reaction of the amine with NITC
Analysis Method:Analysis by HPLC with UV detection at 254 nm
Accuracy:± 25% for all amines except methyl amine
± 35% for methyl amine
Shelf-life:12 months from date of manufacture in freezer
Storage:

Before use: ≤39.2 F (4 C)
After use: Samples can store at ambient temperatures for 14 days
For storage greater than 14 days, store at ≤39.2 F (4 C).

Do not store with food. 

Temperature Effects:No effect on sampling rate between 50 and 86 F (10 and 30 C)
Humidity Effects:No effect on sampling rate from 10 to 80% relative humid-ity (RH)
Wind Velocity Effects:No effect from 5 to 100 cm/sec
Interferences:None found; highly specific for amines
Validation:Partial
Dimensions:3.4 x 1.1 x 0.35 in (8.6 x 2.8 x .89 cm)
Weight:0.38 oz (10.8 g)
Slide Cover:Blue