AirChek XR5000 Button Sampler Pump Kit

Part Number:

210-4121

- +
Overview
Button Sampler Pump Kit includes Button Sampler, standard AirChek® XR5000 Sample Pump, single charger (100-240 V), 3 feet of Tygon tubing, and calibration adapter; requires a 25-mm filter. AirChek XR5000 pumps contain Li-Ion batteries and are subject to special shipping regulations. AirChek XR5000 requires ¼-inch ID tubing.
The Button Sampler, with its porous curved-surface inlet, is designed to improve the collection characteristics of inhalable dust, including bioaerosols for viable or non-viable analysis. Use the Button Sampler for low-level personal or area sampling of inhalable dust.
Attributes
Flow Rate5 to 5000 ml/min
Touch Screen OperationNo
Battery Type (Power)Li-ion
Approx. Run Time20 Hours at 2L/min
ConnectivityNo
PC CompatibleNo
ProgrammableYes
Compensates for Changes in Atmospheric PressureNo
Compensates for Changes in TemperatureYes
Constant FlowYes
Constant Pressure ModeNo
Automatic Flow Rate VerificationNo
Intrinsic SafetyYes
CE MarkedYes
UKCA MarkedYes
 

Performance Profile

Flow Rate4 L/min*
Sampling InletConductive stainless steel
BodyAluminum
Support ScreenStainless steel
ClipStainless steel and nylon
O-ringsPTFE (inlet) and BUNA-N (body)
Filter25 mm**
Inhalable DustGravimetric (GR)
Fungal SporesEpifluorescence microscopy, immunoassay, or polymerase chain reaction (PCR)
MetalsX-ray Fluorescence (XRF)
Tubing1/4-in ID
*The Button Sampler closely follows the ISO 7708/CEN inhalability curve at 4 L/min. This provides optimum sampling.
** A filter pore size of 1.0 µm or larger is recommended for use with the Button Sampler to reduce back pressure.