AirLite 5-pack High Flow Pump Kit

Part Number:

110-100K5

- +
Overview
AirLite® 5-pack High Flow Pump Kit includes 5 pumps with 5 sets of 3 AA alkaline batteries and screwdriver set and filter cassette holders, in a hard-sided case. Low flows from 5 to 500 ml/min require a low flow kit.
The AirLite personal air sampling pump offers a flow range of 5 to 3000 ml/min, economical alkaline battery operation, and over 10-hour run times for sampling gases and vapors and particulates. The pump is ideal for use with sorbent tubes, filter cassettes, and size-selective samplers. The AirLite air sampling pump is not UL Listed for intrinsic safety.
Attributes
Flow Rate5 to 3000 ml/min
Touch Screen OperationNo
Battery Type (Power)Alkaline
Approx. Run Time10+ Hours at 2 L/min
ConnectivityNo
PC CompatibleNo
ProgrammableNo
Compensates for Changes in Atmospheric PressureNo
Compensates for Changes in TemperatureYes
Constant FlowYes
Constant Pressure ModeNo
Automatic Flow Rate VerificationNo
Intrinsic SafetyNo
CE MarkedYes
UKCA MarkedYes
 

Performance Profile

Compensated Flow Range

1000 to 3000 ml/min, ± 5%
(5 to 500 ml/min requires low flow accessories; see Low Flow Kit)
Requires 1/4-inch ID tubing

Flow Control
Patented (U.S. Patent No. 6,741,056) system holds constant flow to ± 5% of the set-point
Compensation Range for a Minimum of 8-hour Operation
1000 ml/min at 30 inches water back pressure
2000 ml/min at 20 inches water back pressure
3000 ml/min at 10 inches water back pressure
 Typical Back Pressure of Sampling Media (inches water)
Flow Rate (L/min)11.522.53
Filter/Pore Size (µm)     
25-mm MCE/0.8691215
25-mm MCE/0.451422
37-mm MCE/0.823456
37-mm PVC/5.011222.5
Compare the information in this table to pump compensation range to determine appropriate applications.
Temperature Range
Operating: 32 to 104 F (0 to 40 C)
Storage: -4 to 113 F (-20 to 45 C)

Protect sample pump from weather when in use outdoors.
Compensated Temperature Range
41 to 122 F (5 to 50 C )
Operating Humidity
0 to 95% non-condensing
Power
Three standard AA alkaline batteries
Rechargeable AA-size 1.2-volt NiMH batteries may be used but will provide approximately half the run time stated for disposable batteries.
LED Functions

Slow flash = normal run
Fast flash = flow fault
Double flash = low battery, approximately one-hour run time remaining
Flash every eight seconds = pump shutdown due to exhausted batteries
Indication of pump shutdown due to exhausted batteries will override all other indicators.
Four flashes every six seconds = pump shutdown due to flow fault

Run Time
(new batteries)
8.5 hours at 1 L/min, 25 inches water back pressure
10 hours at 2.5 L/min, 15 inches water back pressure

Dependent on batteries used. See tables below.
Tubing Requires 1/4-inch ID tubing
Case
Plastic, RFI/EMI-shielded, CE marked
Dimensions
4.6 x 3 x 1.75 in (11.7 x 7.6 x 4.4 cm) includes inlet and belt clip
Weight
(includes batteries)
12 oz (340 gm)
Caution:
Do not operate AirLite in hazardous or explosive locations. AirLite is designed for applications that do not require intrinsic safety.